信息化追溯系统联合器械图谱技术在再生硬式内镜器械中的应用效果

    Application effects of the information traceability system combined with the instrument atlas technology in the regenerative rigid endoscopic instruments

    • 摘要: 目的:评估将信息化追溯系统与器械图谱技术整合并应用于可重复使用的硬式内镜器械再处理中的有效性。方法:采用便利抽样法,进行信息化追溯系统与器械图谱技术整合应用前后比较。对照组为2024年6—8月期间使用常规方法处理的54包(共648件)器械;观察组为2024年9—11月期间使用信息化追溯系统与器械图谱技术整合应用后处理的54包(共648件)器械。比较2组器械的清洗、拆卸和组装合格率,同时评估器械管理评分、处置缺陷发生率以及手术科室的满意度。结果:观察组在清洗、拆卸和组装的合格率高于对照组(P<0.05~ P<0.01);观察组器械管理评分和手术科室满意度更高(P<0.01);观察组器械包处置缺陷总发生率明显低于对照组(P<0.05)。结论:将信息化追溯系统与器械图谱技术整合应用于可重复使用的硬式内镜器械再处理流程,可提升处理质量、提高操作效率、降低缺陷率,并增强手术科室的满意度。

       

      Abstract: Objective: To evaluate the effectiveness of the information traceability system combined with the instrument atlas technology in the application of reprocessing of regenerative rigid endoscopic instruments. Methods: The convenience sampling method was adopted to conduct a comparison before and after the integrated application of the information traceability system and instrument atlas technology.The control group consisted of 54 packages(a total of 648 pieces) of instruments treated with conventional methods from June to August 2024.The observation group consisted of 54 packages(a total of 648 pieces) of instruments treated with the integrated application of information traceability system and instruments atlas technology from September to November 2024.The qualified rates of cleaning,disassembly and assembly of instruments between two groups were compared.Meanwhile,the instrument management score,incidence of disposal defects and satisfaction of the surgical department were evaluated. Results: The qualified rates of cleaning,disassembly and assembly in the observation group were higher than those in control group(P<0.05 to P<0.01).The score of instrument management and satisfaction of surgical department in the observation group were higher than those in control group(P<0.01).The total incidence of instrument package disposal defects in the observation group was significantly lower than that in control group(P<0.05). Conclusions: The information traceability system combined with the instrument atlas technology in the application of reprocessing of regenerative rigid endoscopic instruments can improve the processing quality,increase the operational efficiency,reduce the defect rates and enhance the satisfaction of surgical department.

       

    /

    返回文章
    返回