XE-2100型全自动血液分析仪2种血小板计数方法的差异性比较

    Comparison of two methods of platelet counting by XE-2100 automatic hematology analyzer

    • 摘要: 目的: 观察XE-2100型全自动血液分析仪激光法(PLT-O)和电阻法(PLT-I)计数血小板(PLT)的准确性。方法: 收集高值(PLT > 300×109/L)、中值(80×109/L < PLT ≤ 300×109/L)、低值(PLT ≤ 80×109/L) PLT标本各40份,分别在XE-2100型血液分析仪上检测,观察PLT-O和PLT-I的结果,同时与手工显微镜法结果相比较。结果: XE-2100型血液分析仪计数PLT时,PLT-O和PLT-I结果均与显微镜计数法结果差异均无统计学意义(P > 0.05);对于低值PLT组,PLT-O结果更接近显微镜计数法结果,而PLT-I结果偏低。结论: XE-2100型血液分析仪计数高值和中值PLT组,PLT-O和PLT-I结果一致;而在低值PLT复检时,PLT-O结果具有重要的参考价值。

       

      Abstract: Objective: To observe the accuracy of optical (PLT-O) method and electric impedance (PLT-I) method in counting the platelet by the XE-2100 automatic hematology analyzer. Methods: The platelet samples were collected from the high-value group (40 cases)(PLT > 300×109/L),median-value group (40 cases)(80×109/L < PLT ≤ 300×109/L) and low-value group (40 cases)(PLT ≤ 80×109/L). The platelet counting was conducted using PLT-O method and PLT-I method with the XE-2100 automatic hematology analyzer,and the results were compared with those by manual method. Results: The results obtained by PLT-O and PLT-I methods were consistent with those by manual method in the median-value group and the high-value platelet group (P > 0.05). However,of those lowvalue groups,the results of PLT-O are more consistent with the results of manual method than the one of PLT-I. Conclusions: The counting results by PLT-O and PLT-I methods using the XE-2100 hematology analyzer are consistent in the median-value and the highvalue platelet groups,but results by PLT-O have more important reference value in the re-examination to low-value platelet group.

       

    /

    返回文章
    返回