血糖仪与Olympus AU640型生化分析仪测定血糖的比对分析

    Comparison of handheld glucose meter and Olympus AU640 biochemistry analyzer

    • 摘要: 目的: 探讨临床科室血糖仪检测与生化分析仪检测结果的一致性。方法: 对比25台血糖仪和Olympus AU640型生化分析仪检测(低、中、高)水平的血糖标本结果,分析各水平各血糖仪的偏倚。结果: 25台中1台血糖仪测定高水平血糖标本时偏倚为22.3%,> 20%判为不合格;血糖仪测定低水平血糖标本偏倚为-0.3%~-11.8%,中水平偏倚为-2.4%~7.9%;高水平为-8.9%~7.8%。结论: 血糖仪检测与生化分析仪检测结果一致性较好,检测结果可靠,但还应定期对血糖仪进行质控。

       

      Abstract: Objective: To comprehend the concordance of the test results by the handheld glucose meter and the biochemistry analyzer. Methods: The blood glucose samples (low, medium and high levels) were tested by 25 blood glucose meters and the Olympus AU640 biochemistry analyzer. The results and the bias in testing each level of glucose samples were compared and analyzed. Results: The bias reached 22.3% in testing high glucose specimen by 1 of the 25 glucose meters (> 20% as unqualified). The bias was -0.3% to 11.8% in testing low glucose samples,-2.4% to 7.9% in mid-level glucose samples and -8.9% to 7.8% in high glucose samples by the handheld glucose meters. Conclusions: The concordance of the test results by the handheld glucose meter and the biochemistry analyzer is good. The results are reliable, but quality control on the blood glucose meters should be regularly carried out.

       

    /